Search Results (2 CVEs found)

CVE Vendors Products Updated CVSS v3.1
CVE-2026-4216 1 I-sens 1 Smartlog App 2026-03-17 5.3 Medium
A weakness has been identified in i-SENS SmartLog App up to 2.6.8 on Android. This affects an unknown function of the component air.SmartLog.android. This manipulation causes hard-coded credentials. The attack can only be executed locally. The exploit has been made available to the public and could be used for attacks. The vendor explains: "The function referenced in the report currently exists in our deployed system. It is related to a developer mode used during the configuration process for Bluetooth pairing between the blood glucose meter and the SmartLog application. This function is intended for configuration purposes related to device integration and testing. (...) [I]n a future application update, we plan to review measures to either remove the developer mode function or restrict access to it."
CVE-2017-13993 1 I-sens 1 Smartlog Diabetes Management Software 2025-04-20 N/A
An Uncontrolled Search Path or Element issue was discovered in i-SENS SmartLog Diabetes Management Software, Version 2.4.0 and prior versions. An uncontrolled search path element vulnerability has been identified which could be exploited by placing a specially crafted DLL file in the search path. If the malicious DLL is loaded prior to the valid DLL, an attacker could execute arbitrary code on the system. This vulnerability does not affect the connected blood glucose monitor and would not impact delivery of therapy to the patient.